Instructions for Use

Maintenance

6.1 CHARGING THE BATTERY

The TBI system modules can only be used on battery power, therefore you must ensure that the battery is charged regularly. One battery charge is enough for approximately >200 measurements or 3.5 hours of continuous use.

To charge the battery, place the modules on the MESI LARGE CHARGING PLATE UNIT. During the charging process the light on the module will be flashing yellow. Only when the device is completely charged it will start flashing green. For more information see Chapter 5.6 Multifunctional Button.

If the capacity of the battery is significantly decreased after a certain period of intensive use, the battery is most likely spent and should be replaced. As the device does not contain any parts which can be replaced by a user, contact your local distributor or the manufacturer about replacing the battery.

6.2 CLEANING INSTRUCTIONS

⚠️Read and follow cleaning instructions carefully.

Cleaning the Device

The MESI TOE BLOOD PRESSURE UNIT, MESI TUBELESS CUFF UNIT devices, digit cuffs, probes and cuffs must be carefully cleaned and disinfected after every use and before use with a different patient. The digit cuffs/cuffs and probes can be reused on the same patient. If the digit cuff/cuff or probe is covered with dust and dirt, it can affect the performance of the device. Do not use aggressive cleaning agents, volatile liquids or excessive force when cleaning the device. Don’t dip the digit cuff/cuff and probe in liquids or cleaning agents.

Cleaning – MESI TOE BLOOD PRESSURE UNIT/MESI TUBELESS CUFF UNIT:

  • Before cleaning remove the MESI TOE BLOOD PRESSURE UNIT/ MESI TUBELESS CUFF UNIT from the charging station and discon nect the cuffs and the toe blood pressure cable from the units.

  • Use a soft lint-free cloth dampened with a proper cleaning agent to clean and disinfect the exterior/interior of the MESI TOE BLOOD PRESSURE UNIT/MESI TUBELESS CUFF UNIT.

  • With a dry cloth wipe off all cleaning solution residue.

  • Dry the equipment before next use.

  • Recommended cleaning agents: Hydrogen peroxide (3%), Ethanol (70%), Isopropanol (70%), Mild soap (diluted), Ammonia (diluted), Sodium hypochlorite bleach (diluted)

Cleaning – cuffs/probes:

  • Clean the cuff’s surface by wiping a damp cloth with mild soap water.

  • Do not wash the cuffs or immerse it in water. Also do not use petrol, thinners or similar solvents to clean the cuffs.

⚠️Do not sterilise cuffs. Do not use bleach.

⚠️Wash and rinse temperatures should not exceed 40°C.

⚠️Do not, under any circumstances, immerse the unit or the cable assemblies in liquid.

⚠️Do not use high-temperature sterilisation processes (such as autoclaving).

⚠️Do not use e-beam or gamma radiation sterilisation.

NOTE: Do not wash the cuffs in a washing machine or iron them.

NOTE: In the event of mechanical stress, the device must be calibrated!

6.3 DISINFECTION

MESI TOE BLOOD PRESSURE UNIT/MESI TUBELESS CUFF UNIT:

For disinfection use commercially available disinfectants intended for a professional healthcare environment. Refer to the producer's Instructions for Use.

Approved disinfectants:

  • Isopropyl alcohol 70%

  • Propanol (70-80%)

  • Ethyl hexanal

  • Aldehyde (2-4%)

  • Ethanol (70-80%)

Non-approved disinfectants:

  • Organic solvents

  • Ammonia-based detergent

  • Abrasive cleaning agents

  • 100% alcohol, acetone, chlorine etc.

Disinfection of the Cuff/Probe:

  • Can be disinfected with medical-grade alcohol 70%.

  • After disinfection, the cuff cover should be left air dry.

6.4 PRODUCT LIFE AND STORAGE

If correctly used and maintained the device will have a minimum service life of 5 years. It is possible to store the device in suitable conditions for a maximum of 5 years. While in storage the battery needs to be recharged once every three months to avoid accidentally over-discharging the batteries. When using the device after storage, we recommend subjecting the device to a thorough maintenance check.

The device should be inspected by a qualified service engineer at least every 12 months for the following safety checks:

  • Any mechanical or functional damage on the device and accessories,

  • Device performance in accordance with the Instructions for Use,

  • Legibility of warning labels, and

  • Battery cycle count.