Instructions for Use

Technical specifications

The technical information regarding the MESI TOE BLOOD PRESSURE UNIT and the MESI TUBELESS CUFF UNIT provided within the package is as follows:

3.1 MESI TOE BLOOD PRESSURE UNIT (TBPMD)

3.1.1 DIMENSIONS

Width: 40 mm (1.57 inches)

Depth: 40 mm (1.57 inches)

Height: 150 mm (5.91 inches)

Weight: 244 g

3.1.2 POWER & BATTERY

Power & Battery
Column AColumn B
Battery typeRechargeable lithium-polymer battery (LP602248)
Capacity1240 mAh
AC/DC adaptorFW8030M/05 (FRIWO FOX30-XM)
Input100-240V AC/50-60Hz/600-300mA
Output5V DC/5.0A
Examinations per battery charge> 200
Charge Time for Depleted Batteryapproximately 2 hours

3.1.3 CUFF SIZES – TOE

CUFF SIZES – TOE
Column AColumn B
Medium-sized CuffsDigit cuffs pair - Medium
Dimensions90 x 20 mm
Large-sized CuffsDigit cuffs pair - Large
Dimensions120 x 25 mm

3.2. MESI TUBELESS CUFF UNIT (CUFFMD)

3.2.1 DIMENSIONS

MESI TUBELESS CUFF UNIT (CUFFMD)
Column AColumn B
Width40 mm (1.57 inches)
Depth40 mm (1.57 inches)
Height150 mm (5.91 inches)
Weight286 g

3.2.2 POWER & BATTERY

Power & battery
Column AColumn B
Battery TypeRechargeable lithium-polymer battery (LP602248)
Capacity1240 mAh
AC/DC AdaptorFW8030M/05 (FRIWO FOX30-XM)
Input100-240V AC/50-60Hz/600-300mA
Output5V DC/5.0A
Examinations per Battery Charge> 200

3.2.3 CUFF SIZES – ARM

Cuff sizes - arm
Column AColumn B
Medium-sized CuffsTubeless cuff set - BP - Medium
Circumference22-32 cm
Large-sized CuffsTubeless cuff set - BP - Large
Circumference32-43 cm

3.3 DEVICE SPECIFICATIONS

3.3.1 CLASSIFICATION

Classification
Column AColumn B
Protection against Electric ShockClass II
Medical Device ClassificationClass IIa
Applied PartsType BF applied part
Software ClassificationClass B
RF Emissions (CIPSR 11)Group 1. Class A

3.3.2 OPERATING CONDITIONS

Operating conditions
Column AColumn B
Operating Temperature10° to 40°C
Relative Humidity25 to 85% (no condensation)
Pressure during Operation700 to 1060 hPa
Ingress Protection RatingIP42

⚠️ If the device is used or stored outside the specified environmental parameters, the accuracy specified within the technical specifications of the device is not guaranteed.

3.3.3 MEASUREMENT SPECIFICATIONS

Measurements Using Oscillometry, Volume Plethysmography and Photoplethysmography:

  • Toe-Brachial Pressure Index

  • Systolic blood pressure (arms and toes)

  • Diastolic blood pressure (arms)

  • Heart rate

Measurement Range:

  • Arm pressure: 0 to 299 mmHg

  • Toe pressure: 20 to 250 mmHg

  • Heart rate: 30 to 199 beats per minute

Max Deviation:

  • Pressure: ± 3 mmHg

  • Heart rate: ± 5% of value T

  • oe-Brachial Pressure Index: ± 0.1

Temperature sensors for measuring skin temperature are integrated into both PPG probes. Skin temperature can vary significantly from core body temperature and is shown as additional information.

3.3.4 CONNECTIVITY

Data connectivity with MTABMD (Bluetooth 2.1 + EDR)

Receiving Section

Receiving Section
Column AColumn B
Frequency Range2401.3 MHz – 2480.7 MHz
Bandwidth0.930 MHz

Transmitter

Transmitter
Column AColumn B
Output Power0.5 - 4.5 dBm
Frequency Range2401.3 MHz – 2480.7 MHz
ModulationGFSK

ChatBot