ErrorDescriptionSolution
E2: An anomaly has been detected. The patient may have moved during the measurement process.An anomaly has been detected.Remind the patient to remain still during measurement and repeat the measurement.
E3: The cuff was insufficiently inflated.Inflation insufficiency.Check the placement of the cuffs and repeat the measurement.
E4: An error occurred during heart rate calculation.An error occurred during heart rate calculation.Repeat the measurement. if the error is repeated, the measured value is outside the measurement range of the device.
E6: Inflation error. Check the placement of the cuffs and repeat the measurement.An error occurred during inflation of the cuff.Check the placement of the cuffs and repeat the measurement.
E7: Deflation too fast. The cuff is not attached or is not attached correctly. Place the cuff correctly and repeat the measurement.The cuff was insufficiently deflated.Check the placement of the cuffs and repeat the measurement.
E8: An error occurred during systolic pressure calculation.An error occurred during systolic pressure calculation.Repeat the measurement. if the error is repeated, the measured value is outside the measurement range of the device.
E9: An error occurred during diastolic pressure calculation.An error occurred during diastolic pressure calculation.Repeat the measurement. if the error is repeated, the measured value is outside the measurement range of the device.
E11: A large pressure fluctuation was detected. The patient may have moved.A large pressure fluctuation was detected. The patient may have moved.The results may be incorrect. Repeat the measurement.
E21: MESI TUBLESS CUFF UNIT/MESI TOE BLOOD PRESSURE UNIT not in range.The main unit is out of range (timeout).Bring the main unit closer to the tablet and restart the TBI application.
E21: MESI TUBLESS CUFF UNIT/MESI TOE BLOOD PRESSURE UNIT not in range.Connection between the main unit and MESI mTABLET UNIT is not optimal.Remove any mobile devices in proximity of the main unit or bring the MESI mTABLET UNIT closer to the main unit.
E21: MESI TUBLESS CUFF UNIT/MESI TOE BLOOD PRESSURE UNIT not in range.Depleted battery.Place the main unit next to the charging station.
E22: MESI TUBLESS CUFF MODULE/MESI TOE BLOOD PRESSURE MODULE not paired with MESI mTABLET UNIT.The main unit is not paired to the MESI mTABLET UNIT.Pair the main unit with the MESI mTABLET UNIT. For pairing see Chapter 5.1.5 Pairing.
E23: MESI TUBLESS CUFF MODULE/MESI TOE BLOOD PRESSURE MODULE battery cycle count is high.Battery needs to be replaced.Please contact MESI representative to replace the battery.
E24: Depleted battery. Measurement ended.TBPMD/CUFFMD battery is empty.Place the TBI UNIT to the MESI LARGE CHARGING PLATE UNIT.
N25: Connection to the diagnostic module has been interrupted. Please restart the recording process.Connection to the diagnostic module has been interrupted. Please restart the recording process.Restart the recording process.
N28: Measurement stopped due to low battery. Please recharge the module battery of MESI TUBELESS CUFF MODULE/ MESI TOE BLOOD PRESSURE MODULE and restart the measurement.Measurement stopped due to low battery. Please recharge the module battery of MESI TUBELESS CUFF MODULE/ MESI TOE BLOOD PRESSURE MODULE and restart the measurement.Recharge the module battery of MESI TUBELESS CUFF MODULE/ MESI TOE BLOOD PRESSURE MODULE and restart the measurement.
N29: Connection to the diagnostic module has been interrupted. Please restart the recording process.Connection to the diagnostic module has been interrupted.Please restart the recording process.
E41: No cuff attached to MESI TUBELESS CUFF UNIT.No cuff attached to MESI TUBELESS CUFF UNIT.Attach the cuff to TUBELESS CUFF UNIT.
E42: Please check the cuff position and repeat the measurement.\nIf error is repeated, the measured value is outside the measurement range of the device.Please check the cuff position and repeat the measurement. \nIf error is repeated, the measured value is outside the measurement range of the device.Check the cuff position and repeat the measurement.
N101: The FirstWave™ algorithm was not able to detect any pulse on at least one toe. Please select it manually or confirm the result.No blood flow in toes detected.Select the point of return for blood flow manually or confirm the measurement without re-positioning the cursor.