Instructions for Use

General warnings

8.1 PATIENT INJURY PREVENTION

⚠️Do not place the electrodes on wounds, as this may cause additional injury.

⚠️Before using the device for the first time, read the user manual carefully and follow the recommendations and suggestions.

⚠️This unit is CF classified according to IEC 60601-1. This means that the patient connection is fully isolated and defibrillation protected. The manufacturer can only guarantee protection against defibrillation voltage however, when the original MESI patient cable is used.

⚠️There is no danger when using the MESI ECG UNIT for a pacemaker patient or with simultaneous use of other electrical stimulation equipment. However, the stimulation units should only be used at a sufficient distance from the electrodes. In case of doubt, the patient should be disconnected from the recorder.

⚠️It must be ensured that neither the patient nor the electrodes (including the neutral electrode) come into contact with other persons or conducting objects (even if these are earthed).

⚠️While the electrocardiograph is safety-rated CF for direct cardiac contact, it is not intended to be connected directly to the patient’s heart. Only surface contact with the patient’s skin is intended.

8.2 RECORDING

⚠️The MESI mTABLET ECG users must be adequately trained to use the device. Before the first use of the device, users must carefully read the entire instructions for use and follow the instructions for use of the connected equipment.

⚠️MESI mTABLET ECG complies with EMC regulations for medical products which affords protection against emissions and electrical interference. However special care must be exercised when the unit is used with high frequency equipment.

⚠️Use only accessories and other parts recommended or supplied by MESI. Use of other than recommended or supplied parts may result in injury inaccurate information and/ or damage to the unit.

⚠️This product is not designed for sterile use.

⚠️Do not use this unit in areas where there is any danger of explosion or the presence of flammable gases such as anaesthetic agents.

⚠️Filters may suppress diagnostically relevant portions of the signal, because they limit the transmission range. Therefore, only enable f ilters if necessary.

8.3 MAINTENANCE

⚠️If device is used or stored outside the specified temperature range and air humidity, the accuracy specified within the technical specifications of the device is not guaranteed.

⚠️Before cleaning disconnect from the mains. Due to constant standby mode do not clean device terminals.

⚠️Do not use the device when it is wet. After cleaning the device with a damp cloth, wait for it to dry. Only use the device when it is completely dry.

⚠️Do not, under any circumstances, immerse the unit or cable assemblies in liquid.

⚠️Only use non-aggressive cleaning agents to clean the device. The device may be wiped with a damp cloth.

⚠️Do not use high temperature sterilization processes (such as autoclaving). Do not use e-beam or gamma radiation sterilization.

⚠️Make sure that the device does not come into contact with an electrical current while it is being cleaned.

⚠️Do not dispose of the device as unsorted municipal waste. Prepare it for recycling or separate waste collection in accordance with Directive 2002/96/EC on scrap electrical and electronic equipment (WEEE).

⚠️To prevent electric shock hazard due to current leakage, only use AC/DC power supplies which are compliant with the technical specifications of the device.

⚠️The AC/DC power supply must be connected to an easily accessible socket (the AC/DC power supply also serves for galvanic isolation).

⚠️To prevent electric shock do not disassemble the unit. No serviceable parts inside. Refer servicing to qualified personnel only.

⚠️The use of cables and accessories other than those approved for use by MESI ltd. may negatively affect EMC performance.

⚠️Inspect patient cable for damage and observe the calibration signal before performing a measurement.

8.4 FUNCTIONING OF THE DEVICE

⚠️The device must only be operated using battery power if the earth connection is suspect or if the mains lead is damaged or suspected of being damaged.

⚠️Protect the device from moisture and liquids and extremely high/low temperatures. Also, protect the device from mechanical stress and do not expose it to direct sunlight, as this can cause the device to not function properly.

⚠️Never carry out repairs of any kind yourself. Do not open the device. The device does not contain any parts which can be replaced by a user. Do not alter or adapt the device. If a defect occurs, consult your dealer or distributor.

⚠️MESI mTABLET ECG module must not be used in an oxygen-rich environment.

⚠️The ECG measurement unit is not intended for use in conjunction with high frequency surgical equipment.

⚠️Important information on electromagnetic compatibility (EMC). As the number of electronic devices such as computers and mobile phones in the room increases, medical devices can become sensitive to the electromagnetic influences of other devices. Electromagnetic interference can cause medical devices to malfunction, which can potentially lead to dangerous situations. Furthermore, medical devices must not interfere with other devices. The IEC/EN 606011-2 standard was introduced due to the necessity to establish electromagnetic compatibility (EMC) requirements for the prevention of dangerous situations in the use of medical devices. The standard defines the level of resistance to electromagnetic interference for medical devices. This medical device is compliant with the IEC/ EN 60601-1-2 standard in terms of resistance to electromagnetic interference and electromagnetic emissions. Nevertheless, do not use mobile phones and similar devices which create strong electromagnetic fields in the vicinity of the device. This can cause the device to malfunction, which can potentially cause a dangerous situation.

⚠️Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm (12 inches) to any part of the device including cables specified by the manufacturer. Otherwise degradation of the performance of this equipment could result.

⚠️Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment should be observed to verify that they are operating normally.