Instructions for Use

Maintenance

7.1 CHARGING THE BATTERY

The device can be only used on battery power therefore you must ensure that the battery is charged regularly. One battery charge is sufficient for approximately 100 recordings.

To charge the battery, please place the MESI ECG UNIT on the MESI charging station.

If the capacity of the battery is significantly decreased after a certain period of intensive use, the battery is most likely spent and you should replace it. As the device does not contain any parts which can be replaced by a user, you should contact your dealer or the manufacturer about replacing the battery.

7.2 CLEANING INSTRUCTIONS

It’s recommended to clean the device regularly with a soft, dry or damp cloth. Do not use aggressive cleaning agents, volatile liquids or excessive force when cleaning the device. Do not wash the patient cables or immerse them in water. Also, do not use petrol, thinners or similar solvents to clean the patient cables. To adequately clean or disinfect the electrodes, use a soft lint-free cloth, which has been soaked in a 70% alcohol solution. Afterwards dry the electrodes well.

7.3 PRODUCT LIFE AND STORAGE

If correctly used and maintained the device will have a minimum service life of 5 years. You can store the device in suitable conditions for a maximum of 5 years. While in storage the battery needs to be recharged every 3 months to avoid accidentally over-discharging your batteries. When using the device after storage, we recommend you to subject the device to a thorough maintenance check.

The device should be inspected by a qualified service engineer at least every 12 months for the following safety checks:

  • any mechanical or functional damage on device and accessories,

  • performance of device in accordance to instructions for use,

  • legibility of warning label,

  • battery cycle count.