Instructions for Use

Errors

Errors
Column AColumn B
ECG measurement unit was not found. ERROR: TimeoutYour ECG measurement unit is out of range. Please bring MESI mTABLET UNIT closer to your ECG measurement unit and restart the ECG application.
Please bring MESI mTABLET UNIT closer to MESI ECG UNIT. ERROR: Bad connectionWhen performing measurement and the connection between MESI mTABLET UNIT and MESI ECG UNIT is not optimal (i.e. you are performing measurement close to mobile devices with wireless connections) please bring your MESI mTABLET UNIT closer to MESI ECG UNIT or remove any mobile devices in proximity of MESI ECG UNIT.
Please bring MESI mTABLET UNIT closer to measuring module and repeat the measurement. ERROR: Bad connectionIf the measurement was interrupted because of bad connection, please bring your MESI mTABLET UNIT closer to MESI ECG UNIT or remove any mobile devices in proximity of MESI ECG UNIT. After this please repeat the measuring process.
MESI ECG UNIT not working. ERROR: ECG inoperablePlease contact the manufacturer or the local distributor.
High battery cycle ERROR: Replace batteryBattery will need to be replaced soon, please contact the manufacturer or the local distributor for replacement.
Problem with calibration signal ERROR: Failed calibrationInitial calibration signal problem. Please reset the MESI ECG UNIT by pressing and holding the multifunctional button for about 10 seconds or until it changes colour to red. If the problem continues, please contact the manufacturer or the local distributor.