General warnings

6.1 MEASUREMENT PROCEDURE

⚠️Do not use this unit in areas with a presence of flammable gasses such as anaesthetic agents or in an oxygen-rich environment.

⚠️Monitor the patient closely during the measurement procedure. If there is any reason for concern, stop the test.

⚠️Important information on electromagnetic compatibility (EMC). As the number of electronic devices such as computers and mobile phones in the roomrooms increases, medical devices can become sensitive to the electromagnetic influences of other devices. Electromagnetic interference can cause medical devices to malfunction, which can potentially lead to dangerous situations. Furthermore, medical devices must not interfere with other devices. The IEC/EN 60601-1-2 standard was introduced due to the necessity to establish electromagnetic compatibility (EMC) requirements for the prevention of dangerous situations in the use of medical devices. The standard defines the level of resistance to electromagnetic interference for medical devices. This medical device is compliant with the IEC/EN 60601-1-2 standard in terms of resistance to electromagnetic interference and electromagnetic emissions. Nevertheless, do not use mobile phones and similar devices that create strong electromagnetic fields in the vicinity of the device. This can cause the device to malfunction and potentially cause a dangerous situation.

⚠️Do not operate in proximity of a steam kettle or any equipment that can generate moisture.

⚠️Do not operate in proximity of a fireplace, a or radiant heater or any equipment that can generate heat.

⚠️The cables and accessories may negatively affect the EMC performance. While operated, the device should not be closer than 30 cm (12 inches) from another medical device.

⚠️Do not use near devices that produce microwave radiation as this might interfere with the device operation.

⚠️The contact thermometer probe is designed exclusively for measuring temperature on intact skin. If measuring orally, ensure to use a protective cover.

⚠️Before using the device for the first time, read the Instructions for Use carefully and follow the recommendations.

⚠️The MESI mTABLET BP device users must be adequately educated to use the device. The education must be performed by the trained MESI representative. Before the first use of the device, users must read the entire Instructions for Use carefully and follow the instructions for the use of the connected equipment.

⚠️The MESI mTABLET BP is intended for blood pressure measurement on upper arm.

⚠️Do not use the device when it is wet. After cleaning the device with a damp cloth, wait for it to dry. Only use the device when it is completely dry.

⚠️The MESI mTABLET BP is not intended for use in conjunction with higher frequency surgical equipment

⚠️The AC/DC power supply must be connected to an easily accessible socket (the AC/DC power supply also serves for galvanic isolation).

⚠️When repeating the Blood Pressure measurement for several times a slight pain may appear at the measurement location. Other effects are excluded.

⚠️The cables and accessories may negatively affect the EMC performance. The device while operated should not be stacked closer than 30 cm from another medical device.

⚠️This product is not designed for sterile use.

⚠️Do not use this unit in areas where there in the presence of flammable gases such as anaesthetic agents or in an oxygen-rich environment

⚠️Do not place the sensor on wounds, as this may cause additional injury.

6.2 MAINTENANCE

⚠️Do not dispose of the device as unsorted municipal waste. Prepare it for recycling or separate waste collection in accordance with Directive 2002/96/EC on scrap electrical and electronic equipment (WEEE).

⚠️Only use non-aggressive cleaning agents for cleaning the device. and Mmake sure that the device does not come in contact with an electrical current while it is being cleaned. The device may be wiped with a damp cloth.

⚠️Protect the device from moisture, liquids, extreme temperatures, mechanical stress, and exposure to direct sunlight. If the device is used or stored outside the specified environmental parameters, the parameters specified within the technical specifications of the device are not guaranteed.

⚠️Do not use high-temperature sterilisation processes (such as autoclaving). Do not use e-beam or gamma radiation sterilisation.

⚠️Only perform routine cleaning and maintenance procedures specifically described in these Instructions for Use.

⚠️During transportation or storage, the temperature of the control device may have decreased or increased outside operating temperatures. Do not use the device until it has been at room temperature from 10 °C (50 °F) to 30 °C (86 °F) for at least two hours. This time is required for the thermometer unit to reach its normal recommended operating temperature.

⚠️Never carry out repairs of any kind yourself. Do not open the device. The device does not contain any parts which can be replaced by the user. Do not alter or adapt the device. If a defect occurs, consult your dealer or distributor.

⚠️Do not position the ME equipment so that it is difficult to operate the disconnection device.

⚠️AC/DC power supply unit is considered a disconnecting device from the mains.

⚠️Before cleaning disconnect from the mains. Due to constant standby mode of the device do not clean terminals. 

⚠️To prevent electric shock hazard due to leakage current, only use AC/DC power supplies which are compliant with the technical specifications of the device.

⚠️Only use non-aggressive cleaning agents to clean the device. The device may be wiped with a damp cloth.

⚠️Be careful when wrapping the cuffs. Do not wrap them too tightly

⚠️Make sure that the device does not come into contact with an electrical current while it is being cleaned.

⚠️Read and follow cleaning instructions carefully.

6.3 FUNCTIONING OF THE DEVICE

⚠️The device may only be used by professional medical personnel. The device is class A equipment; it can cause radio interference or even prevent nearby devices from functioning. It may be necessary to reposition the MESI mTABLET THERMO device or protect the room containing the device from electromagnetic radiation.

⚠️The product should not be used if any of its parts are broken, worn, missing, incomplete, distorted, or contaminated.

⚠️It is mandatory to have the device calibrated once per year to ensure the correct functioning and accuracy. Contact your dealer or the manufacturer about calibrating the device.

⚠️Use only accessories and other parts recommended or supplied by MESI. Use of other than recommended or supplied parts may result in injury inaccurate information and/ or damage to the unit. Follow the instructions that are provided with the specific accessory.

⚠️To prevent electric shock or damage to the device do not disassemble the unit. The device does not contain any parts which can be replaced or serviced by a user. If a defect occurs, consult your dealer or distributor

⚠️Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm (12 inches) to any part of the MESI mTABLET SPO2 including cables specified by the manufacturer. Otherwise the degradation of the performance of this equipment could result.

⚠️Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment should be observed to verify that they are operating normally.

⚠️Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.

6.4 PATIENT INJURY PREVENTION

⚠️Consecutively carrying out too frequent measurements may result in injury to the patient!

⚠️In case of the presence of intravenous cannulas or arteriovenous (A-V) fistulas in the limbs, the cuffs and measurement can cause injury to the limb.

⚠️If the patient has had breast surgery, do not place the arm cuff on the side which has been operated on.

⚠️Do not use the device on a patient while they are connected to a vital sign monitor. Pressurization of the cuff can temporarily cause loss of function of simultaneously used monitoring ME equipment on the same limb.

⚠️Check the pressure in the cuff several times during measurement. If the cuff puts pressure on the limb for too long, it can impair the blood flow.

⚠️Be careful when folding the cuffs. Do not fold them too tightly. Follow measuring instructions.