Instructions for Use

Errors

Errors
ErrorDescriptionSolution
N21: The THERMO module is not in range.The MESI THERMO UNIT is out of range.Bring the MESI THERMO UNIT closer to the tablet and restart the Temp application.
N21: The THERMO module is not in range.The connection between the MESI THERMO UNIT and the MESI mTABLET UNIT is not optimal.Remove any mobile devices in the proximity of the MESI THERMO UNIT or bring the MESI mTABLET UNIT closer to the MESI THERMO UNIT.
N22: The THERMO module is not paired with the MESI mTABLET UNIT.The MESI THERMO UNIT is not paired to the tablet.Pair the MESI THERMO UNIT to the MESI mTABLET UNIT. For more information, see chapter 4.2.5 – Pairing.
N30: Depleted battery. Measurement ended.Please disconnect the THERMO module from the power supply and restart the Temp application.Dismantle the MESI THERMO UNIT from the MESI LARGE CHARGING PLATE (CS4SYS) or MESI SMALL CHARGING PLATE (CS2SYS).
N30: Depleted battery. Measurement ended.An error occurred on the THERMO module, or there might be a communication problem between the module and the MESI mTABLET UNIT.Restart the Temp application and the recording process.
N25: Connection to the diagnostic module has been interrupted. Please restart the recording process.The connection between the MESI mTABLET UNIT and the MESI THERMO UNIT has been lost, and the measurement process stopped.Reconnect the MESI THERMO UNIT in the settings on the MESI mTABLET and perform the measurement again.
N27: Measurement was stopped because the stop button was pressed on the diagnostic module.Pressing the status button on the bottom of the IR probe stops the measurement.De-press the status button and restart the measurement process.
N141: The MESI THERMO module is operating outside the allowed temperature range.The measured temperature is below 32 °C or above 43 °C, which may affect the accuracy of the measurement.Ensure the device is used correctly and consider seeking immediate medical attention.
N142: Ensure the device is used correctly and consider seeking immediate medical attention.The surrounding environment’s temperature is below 10 °C or above 40 °C when using the contact probe.Move the MESI THERMO contact probe to an environment between 10 °C and 40 °C.
N143: An error occurred. Please check the THERMO module and try again.A non-specified error occurred; the measurement can’t be performed.Ensure the thermometer is used correctly and repeat the measurement.
N144: The surrounding environment’s temperature is outside the allowed range of THERMO module.The surrounding environment’s temperature is below 15 °C or above 40 °C when using the infrared probe.Move the MESI THERMO infrared probe to an environment between 15 °C and 40 °C.