Instructions for Use

Troubleshooting

Troubleshooting
Column ANew ColumnColumn B
A continuously flashing purple or red indicator on the MESI THERMO UNIT.Incorrect state of the diagnostic module.Restart the MESI THERMO UNIT by pressing and holding the button at the top of the unit. The light should start flashing.
No light shows when the multifunctional button at the top of the MESI THERMO UNIT is pressed.Depleted battery.Place the MESI THERMO UNIT onto the charging station and leave it to charge for at least half an hour before retrying.
Temperature measurement will not start even though the MESI THERMO UNIT is paired and next to MESI mTABLET UNIT.Possibility of electromagnetic interference.Make sure that there are no devices nearby that produce electromagnetic interference. These devices can interfere with the basic functions of the MESI mTABLET UNIT; this can potentially lead to dangerous situations.

NOTE: If the problem continues, contact the manufacturer or the local distributor for further assistance.