Instructions for Use

Maintenance

5.1 CHARGING THE BATTERY

The device can only be used on battery power. Therefore, the battery must be charged regularly. One battery charge is sufficient for approximately 2500 measurements or 19 hours of continuous use.

To charge the battery, place the MESI THERMO UNIT on the MESI LARGE CHARGING PLATE (CS4SYS) or MESI SMALL CHARGING PLATE (CS2SYS). During the charging process, the light on the module will flash orange. Only when the device is completely charged will it start flashing green. For more information, see chapter 4.7 – Multifunctional buttons.

If the capacity of the battery is significantly decreased after a period of intensive use, the battery is most likely spent and should be replaced. As the device does not contain any parts that can be replaced by a user, contact your local distributor or the manufacturer for battery replacement.

5.2. CLEANING AND DISINFECTION

⚠️Before cleaning, disconnect from the mains. Due to constant standby mode of the device, do not clean the terminals as this may cause electrical shock.

⚠️Only perform routine cleaning and maintenance procedures specifically described in these Instructions for Use

MESI THERMOMETER UNIT

Do not use aggressive cleaning agents, volatile liquids or excessive force when cleaning the device. Do not dip the device or accessories in liquids or cleaning agents.

⚠️Do not use aggressive cleaning agents, volatile liquids or excessive force when cleaning the device. Don’t dip the sensor in liquids or cleaning agents.

MESI PULSE OXIMETER UNIT

  • Before cleaning, remove the MESI THERMO UNIT from the charging station and disconnect the sensor from the MESI THERMO UNIT.

  • Use a soft lint-free cloth dampened with a proper cleaning agent to clean and disinfect the exterior and interior of both the IR and contact probes.

  • Wipe off all residues of the cleaning solution with a dry cloth.

  • Dry the equipment before next use.

Recommended cleaning agents:

  • Hydrogen peroxide (3%)

  • Ethanol (70%)

  • Isopropanol (70%)

  • Mild soap (diluted)

  • Ammonia (diluted)

  • Sodium hypochlorite bleach (diluted)

Not approved cleaning agents:

  • Organic solvents

  • Ammonia-based detergent

  • Abrasive cleaning agents

  • 100% alcohol, acetone, chlorine etc.

5.3 PRODUCT LIFE AND STORAGE

If correctly used and maintained, the device will have a minimum service life of 5 years. It is possible to store the device in suitable conditions for a maximum of 5 years. While in storage, the battery needs to be recharged once a month to avoid accidental over-discharge. When using the device after storage, we recommend a thorough maintenance check.

  • any mechanical or functional damage on device and accessories,

  • performance of device in accordance to instructions for use,

  • legibility of labels,

  • battery cycle count.

5.3.1 TRANSPORT & STORAGE CONDITIONS

TRANSPORT & STORAGE CONDITIONS
Column AColumn B
Temperature-25° to 70°C
Relative humidity15 to 90% (no condensation)
Atmospheric pressure500 to 1060 hPa (mbar)

The device should be inspected by a qualified service engineer at least every 12 months for the following safety checks:

  • Any mechanical or functional damage on the device and accessories

  • Performance of device in accordance with the Instructions for Use

  • Legibility of warning label

  • Battery cycle count