3.1 MESI THERMOMETER UNIT (THERMOMD)

3.1.1 Infra-red probe

3.1.1.1 Dimensions

Width: 40 mm (1.57 inches)

Depth: 41.6 mm (1.57 inches)

Height: 153 mm (7,05 inches)

Weight: 213 g

3.1.1.2 Power & Battery
Power & Battery
Column AColumn B
Battery typeRechargeable Lithium-Polymer battery
Capacity2x620 mAh
AC/DC adaptorNEO030.0-I-X-05
Input100-240 V AC / 50-60 Hz / 600-300 mA
Output5V DC / 5.0 A
Number of measurements per battery charge>2500
Continuous use19h
Charge time for depleted batteryapprox. 2h
Number of discharge cycles or years after which a rechargeable battery needs to be replaced500 cycles

3.1.2 Contact probe

3.1.2.1 Dimensions
Contact probe - Dimensions
Column AColumn B
Fi10 mm (0.39 inches) X 9 mm (0.35 inches)
Length118 mm (4.65 inches)
Weight7.5 g
3.1.2.2 Power & Battery
Contact probe - Power and battery
Column AColumn B
Battery typeRechargeable Lithium-Polymer battery
Capacity15 mAh
Wireless charging from IRT
Number of measurements per battery charge>120
Continuous use2h
Charge time for depleted batteryapprox. 2h
Number of discharge cycles or years after which a rechargeable battery needs to be replaced500 cycles

⚠️To prevent electric shock hazards due to current leakage, only use AC/DC power supplies that are compliant with the technical specifications of the device. The AC/DC power supply must be connected to an easily accessible socket (the AC/ DC power supply also serves as galvanic isolation).

3.2 DEVICE SPECIFICATIONS

3.2.1 CLASSIFICATION

Classification
Column AColumn B
Protection against electric shock Class II
Medical device classificationClass IIa
Applied partsType BF Applied part
Software classificationClass B
RF emissions (CIPSR 11)Group 1. Class A

3.2.2 OPERATING CONDITIONS

⚠️Thermometers have a specific temperature range set by the manufacturer. Using them outside this range can cause inaccurate readings.

3.2.2.1 Infra-red probe
OPERATING CONDITIONS - Infra-red probe
Column AColumn B
Operating temperature15° to 40°C
Relative humidity15 to 90% (no condensation)
Pressure during operation700 to 1060 hPa
Ingress protection ratingIP42

NOTE: The probe is protected against the ingress of solid objects larger than 1 mm and against vertically falling water drops when the enclosure is tilted up to 15 degrees.

3.2.2.2 Contact probe
OPERATING CONDITIONS - Contact probe
Column AColumn B
Operating temperature10 to 40 °C
Relative humidity15 to 90% (no condensation)
Pressure during operation700 to 1060 hPa
Ingress protection ratingIP66

NOTE: The probe is dust-tight and protected against powerful water jets.

⚠️If the device is used or stored outside the specified environmental parameters, the accuracy specified within the technical specifications of the device is not guaranteed.

3.2.2 MEASUREMENT SPECIFICATIONS

3.2.3.1 Infra-red probe
Measurement specifications - Infra-red probe
Column AColumn B
Distance from measuring point3 cm
Measurement rangeSurface temperature: 34 to 42 °C Body temperature: 36,4 to 43,8 °C
Extended measurement rangeSurface temperature: 28,6 to 41,2 °C Body temperature: 32 to 43 °C
Accuracy for measurement range± 0,3 °C
Accuracy for extended measurement range± 0,4 °C
3.2.3.2 Contact probe
Measurement specifications - Contact probe
Column AColumn B
Measurement rangeSurface/body temperature: 34 to 42 °C
Extended measurement rangeSurface/body temperature: 32 to 43 °C
Accuracy for measurement range± 0,1 °C
Accuracy for extended measurement range± 0,1 °C

3.2.4 CONNECTIVITY

3.2.4.1 Infra-red probe

Data connectivity with MTABMD (Bluetooth 2.1 + EDR)

Connectivity - Infra-red probe
Receiving sectionColumn B
Frequency range2402 MHz – 2480 MHz
Bandwidth1 MHz
Connectivity - Infra-red probe
TransmitterColumn B
Output power0 dBm
Frequency range2402 MHz – 2480 MHz
ModulationGFSK
3.2.4.2 Contact probe BT low energy
Contact probe BT low energy
Column AColumn B
Frequency range2402 MHz – 2480 MHz
2402 MHz – 2480 MHz1 MHz
1 MHz
Column AColumn B
Output power0 – 8 dBm
Frequency range2402 MHz – 2480 MHz
ModulationGFSK

3.3 CALIBRATION AND CALIBRATION CHECK

The calibration of the MESI mTABLET THERMO module shall be performed yearly by the manufacturer or certified personnel.