Instructions for Use

Troubleshooting

Troubleshooting
Column ANew ColumnColumn B
A continuously flashing purple or red indicator on the MESI SPIROMETER UNIT.Incorrect state of the diagnostic module.Press and hold the button on top of the MESI SPIROMETER UNIT for 15 seconds to restart it. The light should start flashing.
No light shows when the multifunctional button on top of the MESI SPIROMETER UNIT is pressed.Depleted battery.Place the MESI SPIROMETER UNIT onto the charging station and leave it to charge for at least half an hour before retrying.
The spirometry measurement will not start although the MESI SPIROMETER UNIT is paired with and next to MESI mTABLET UNIT.Possibility of electromagnetic interference.Make sure that there aren’t any devices nearby that produce electromagnetic interference. These devices can interfere with the basic functions of the MESI mTABLET UNIT, which can potentially lead to dangerous situations.

NOTE: If the problem persists, contact the manufacturer or local distributor for further assistance.