Instructions for Use

Errors

Errors
ErrorDescriptionSolution
N21: MESI SPIRO MODULE not in range.The MESI SPIROMETER UNIT is out of range.Bring the MESI SPIROMETER UNIT closer to the MESI mTABLET UNIT and restart the SPIROMETRY application.
N21: MESI SPIRO MODULE not in range.The connection between the MESI SPIROMETER UNIT and the MESI mTABLET UNIT is not optimal.Remove any mobile devices in proximity of the MESI SPIROMETER UNIT or bring the MESI mTABLET UNIT closer to the MESI SPIROMETER UNIT.
N22: MESI SPIRO MODULE not paired with MESI mTABLET UNIT.The MESI SPIROMETER UNIT is not paired to the tablet.Pair the MESI SPIROMETER UNIT to the MESI mTABLET UNIT. For pairing, see chapter 5.1.5 Pairing.
N23: MESI SPIRO MODULE battery cycle count is high.The battery needs replacing.Please contact your MESI representative to replace the battery.
E24: Depleted battery. Measurement ended.The battery of the MESI SPIROMETER UNIT is empty.Place the MESI SPIROMETER UNIT onto the MESI LARGE CHARGING PLATE UNIT.
N61: Please disconnect the MESI SPIRO MODULE from the power supply and restart the SPIROMETRY application.The MESI SPIROMETER UNIT is placed on the MESI LARGE CHARGING PLATE UNIT.Take the MESI SPIROMETER UNIT off the MESI LARGE CHARGING PLATE.
N62: Connection to diagnostic module has been interrupted. Please restart the recording process.A SPIRO module error has occurred, or there might be a communication problem between the module and the MESI mTABLET.Restart the SPIROMETRY application and the recording process.