Instructions for Use

Quick measuring guide

NOTE: Before using the device for the first time, read the Instructions for Use carefully and follow their recommendations and suggestions. This chapter only contains short instructions for the use of the MESI mTABLET SPIRO. For detailed description of the individual functions of the device, see chapter 5.3.1 Step-bystep SPIRO measurement.

4.1 PREPARATION FOR MEASUREMENT

⚠️The MESI SPIRO MODULE is part of the MESI mTABLET SPIRO system. Before starting a measurement, please make sure that you are familiar with all devices that are part of the system as well as with their instructions. The MESI mTABLET SPIRO is comprised of the MESI mTABLET (MTABSYSW), the MESI SPIRO MODULE (SPIROSYS) and the MESI LARGE CHARGING PLATE (CS4SYS).

4.1.1 PAIRING WITH A MESI mTABLET UNIT

Before any measurements can be performed, the MESI SPIROMETER UNIT module needs to be paired to the MESI mTABLET UNIT. For detailed instructions, please follow the instructions in chapter 5.1.5 Pairing.

4.1.2 MOUNTING THE MOUTHPIECE TO THE UNIT

The mouthpieces are disconnected during storage and transport, and are individually packed in separate sealed bags in the transportation box.

Before using the spirometer, the user shall mount the mouthpiece onto the MESI SPIROMETER UNIT.

Connect the mouthpiece to the top side port on the MESI SPIROMETER UNIT. Position, twist and lock the mouthpiece from the grey dot to green in order to assure secure mounting of the mouthpiece.

MESI mTABLET SPIRO - Mounting the mouthpiece to the unit

⚠️Before starting actual measurements or calibration, it is recommended to allow the spirometer device to have a brief warm-up period of a minimum of 5 minutes to reach thermal equilibrium.

4.2 PERFORMING FEVC MEASUREMENT

NOTE: For additional information, see the MESI mTABLET Instructions for Use, chapter 9 MANAGING USERS.

Step 1

On the MESI mTABLET UNIT, select an existing patient (1) or add a new one (2).

MESI mTABLET SPIRO - Performing FEVC measurement - Step 1

Step 2

After selecting the patient, select one of the SPIRO measurements in the application menu.

MESI mTABLET SPIRO - Performing FEVC measurement - Step 2

Step 3

After pressing CONTINUE, enter the patient data (height, weight) and select the appropriate prediction model from the dropdown menu.

MESI mTABLET SPIRO - Performing FEVC measurement - Step 3

Step 4

After entering the measuring screen, explain the measurement and prepare the patient for the maneuver before starting the actual measurement. The patient shall breathe normally at a relaxed rate.

MESI mTABLET SPIRO - Performing FEVC measurement - Step 4

Step 5

After pressing the START FEVC button, the device goes into standby mode and starts measuring when the patient starts blowing.

Before the blow, the patient inspires long and slowly. The patient expires fast and strongly into the mouthpiece, blowing for at least 6 seconds (till the animation ends).

Finish the maneuver by pressing the END RECORDING button.

MESI mTABLET SPIRO - Performing FEVC measurement - Step 5

Step 6

After pressing END RECORDING, the patient can remove the mouthpiece from their mouth. The user can review the maneuver spirograph and values.

MESI mTABLET SPIRO - Performing FEVC measurement - Step 6

Step 7

Performing the maneuver can be repeated by pressing the NEW RECORDING (1) button. By pressing the FINISH (2) button, the measurement can be completed. It is advised that the patient performs at least three representative maneuvers, the best of which will automatically be selected as the result.*

MESI mTABLET SPIRO - Performing FEVC measurement - Step 7

Step 8

After pressing the FINISH button, select the patient’s level of cooperation from the dropdown menu and add comments to the measurements.

MESI mTABLET SPIRO - Performing FEVC measurement - Step 8

*According to the ATS/ERS 2019 recommendations.

4.3 RESULTS

Once the measurement has been taken, the system will automatically switch to the results screen. By scrolling the results page, additional SPIRO measurement parameters can be viewed. The following actions are listed on the top the navigation menu:

MESI mTABLET SPIRO - Results

NOTE: For more information about the result screen, see chapter 5.4 REVIEWING SPIRO MEASUREMENT.