Instructions for Use

Troubleshooting

Troubleshooting
Column ANew ColumnColumn B
Unexpected result.Incorrect cuff placement.Reread the instructions for use and place the cuffs correctly.
Unexpected result.Patient moving during measurement.Repeat the measurement process.
Unexpected result.Wrong cuff size used.Use cuffs of the correct size.
Unexpected result.Possible air leakage.Check the cuffs, the air tubes and the connectors and replace them if necessary. If you cannot fix the issue yourself, consult your dealer or the manufacturer.
Audible stretching of the fastening tapeIncorrect cuff placement.Reread the instructions for use and place the cuffs correctly.
Audible stretching of the fastening tapeWrong cuff size used.Use cuffs of the correct size.
Continuously flashing purple or red indicator on CUFFMDIncorrect state of the diagnostic modulePress and hold the button on top of the MESI TUBELESS CUFF UNIT (CUFFMD) for 15 seconds to restart it.
No light indication when pressing the CUFFMD multifunctional button.Battery is emptyPlace the MESI TUBELESS CUFF UNIT on the charging station and leave it to charge for at least half an hour before retrying.
The cuffs do not inflate. Hissing noises.Possible air leakage.Check the cuffs, the air tubes and the connectors and replace them if necessary.

NOTE: If the problem continues, contact the manufacturer or the local distributor for further assistance.