Instructions for Use

Quick measuring guide

NOTE: Before using the device for the first time, read the Instructions for Use carefully and follow the recommendations and suggestions. This Chapter only includes short instructions for the use of the MESI mTABLET ABI. For detailed description of short instructions on the device, see Chapter 5 DETAILED INSTRUCTIONS.

NOTE: When performing the Ankle-Brachial Index measurement the patient must be in supine position and remain still.

NOTE: The MESI mTABLET ABI is intended for use in professional environment, where measurements must be carried out by adequately trained medical personnel. The MESI mTABLET ABI is not intended for home use.

NOTE: The MESI mTABLET ABI may be used on pregnant women.

NOTE: The MESI mTABLET ABI is not intended for use on new-borns or children under the age of 10 years.

NOTE: In case of the presence of intravenous cannulas or arteriovenous (AV) fistulas, the cuffs and measurement can cause injury to the limb.

4.1 PREPARATION FOR MEASUREMENT

⚠️MESI TUBELESS CUFF MODULES are a part of the MESI mTABLET ABI system. Before starting a measurement be sure that you are familiar with all devices and their instructions which are part of the system.

4.1.1 PAIRING WITH A MESI mTABLET UNIT

Before any measurements can be performed all MESI TUBELESS CUFF UNIT devices need to be paired to the MESI mTABLET UNIT. For detailed instructions please follow instructions in Chapter 5.1.5 Pairing.

4.1.2 ASSEMBLY OF THE MESI TUBELESS CUFF UNIT

During shipping and transportation, the conical cuffs are disconnected from MESI TUBELESS CUFF UNITS. Before first use connect the cuffs to the port on the MESI TUBELESS CUFF UNIT as shown on the image below. Ensure that all cuffs are securely attached.

ASSEMBLY OF THE MESI TUBELESS CUFF UNIT

4.1.3 PATIENT PREPARATION

The patient must be in supine position and remain still and quiet. One thin layer of clothing is allowed.

Step 1 Choosing the right colour of the cuff

Select the appropriate cuff, depending on the description and the colour of the cuff:

Choosing the right colour of the cuff
POSITIONDESCRIPTION on the cuffCOLOR on the cuff
Right armRIGHT ARMRED
Left armLEFT ARMYELLOW
Right ankleRIGHT ANKLEBLACK
Left ankleLEFT ANKLEGREEN
ABI cuffs

NOTE: The cuffs can be connected to either of the MESI TUBELESS CUFF UNIT. The colour, size and positioning will be automatically detected by the MESI TUBELESS CUFF UNIT.

Step 2 Place the cuffs to the appropriate arm/leg

ARM:

  • Position the cuff 1-2 cm above the elbow joint. Align the label Artery with the Artery on inner side of an arm.

  • Place the cuff so that there is two fingers’ width of room between the limb and the cuff.

  • Check that you have chosen the correct size using the INDEX marking and the OK area of the cuff.

ANKLE:

  • Position the cuff 2-3 cm above the ankle. Make sure that the label Ankle points towards inner side of an ankle.

  • Place the cuff so that there is two fingers’ width of room between the limb and the cuff.

  • Check that you have chosen the correct size using the INDEX marking and the OK area on the cuff.

NOTE: Observe instructions for use that are provided with the cuff.

4.1.4 PERFORMING ABI MEASUREMENT

Step 1

On the MESI mTABLET UNIT select an existing patient (1) or add a new one (2).

Patient preparation - MESI mTABLET ABI

Step 2

After selecting the patient, select the ABI measurement in the application menu.

Patient preparation - MESI mTABLET ABI - Step 2

Step 3

When starting the ABI application for the first time the default settings can be set by selecting between a 3-cuff or 4-cuff ABI measurement (when selecting a 3-cuff measurement, please select right or left arm) and confirming. These settings can be changed at any time by pressing "Options" .

Patient preparation - MESI mTABLET ABI - Step 3

Step 4

Observe the position indication on the cuffs and place them on appropriate arm or leg. Then press START (1) and wait until the measurement is completed (2).

Patient preparation - MESI mTABLET ABI - Step 4 (1)Patient preparation - MESI mTABLET ABI - Step 4 (2)

4.2 RESULTS

Once the measurement has been taken the system will automatically switch to the results page. By scrolling on the results screen the ABI, oscillation graph, pulse waveform and SmartArm™ selection can be reviewed.

On the top the navigation menu provides the following actions:

Results - MESI mTABLET ABI

NOTE: For more information about the result screen see Chapter 5.4 REVIEWING ABI MEASUREMENTS