6.1 CHARGING THE BATTERY

The device can be only used on battery power therefore it has to be ensured that the battery is charged regularly. One battery charge is enough for approximately 200 measurements.

To charge the battery, place the MESI TUBELESS CUFF UNIT on the MESI LARGE CHARGING PLATE UNIT. During the charging process the light on the module will be flashing yellow. Only when the device is completely charged it will start flashing green. For more information see chapter 5.5 Multifunctional button.

If the capacity of the battery is significantly decreased after a certain period of intensive use, the battery is most likely spent, and it should be replaced. As the device does not contain any parts which can be replaced by a user, contact your local distributor or the manufacturer about replacing the battery.

6.2 CLEANING INSTRUCTIONS

⚠️Read and follow cleaning instructions carefully.

Cleaning – MESI TUBELESS CUFF UNIT:

  • Before cleaning remove the MESI TUBELESS CUFF UNIT from the charging station and disconnect the cuff from the MESI TUBELESS CUFF UNIT.

  • Use a soft lint-free cloth dampened with a proper cleaning agent to clean and disinfect the exterior/interior of the unit/.

  • With a dry cloth wipe off all residues of the cleaning solution.

  • Dry the equipment before next use.

  • Recommended cleaning agents: Hydrogen peroxide (3%), Ethanol (70%), Isopropanol (70%), Mild soap (diluted), Ammonia (diluted), Sodium hypochlorite bleach (diluted)

Cleaning – cuff:

  • Clean the cuff’s surface by wiping a damp cloth with mild soap water.

  • Do not wash the cuffs or immerse it in water. Also do not use petrol, thinners or similar solvents to clean the cuffs.

⚠️Do not sterilise cuffs. Do not use bleach.

⚠️Wash and rinse temperatures should not exceed 40°C.

NOTE: Do not wash the cuffs in a washing machine or iron them.

NOTE: The device must be calibrated in the event of mechanical stress!

6.3 DISINFECTION

Disinfection – MESI TUBELESS CUFF UNIT:

Use commercially available disinfectants intended for the professional healthcare environment. Refer to the instruction for use of the producer.

Approved disinfectants:

  • Isopropyl alcohol 70%

  • Propanol (70-80%)

  • Ethyl hexanal

  • Aldehyde (2-4%)

  • Ethanol (70-80%)

Non-approved disinfectants:

  • Organic solvents

  • Ammonia-based detergent

  • Abrasive cleaning agents

  • 100% alcohol, acetone, chlorine etc.

Disinfection – cuff:

  • Can be disinfected with medical-grade alcohol 70%.

  • After disinfection, the cuff cover should be left air dry.

6.4 PRODUCT LIFE AND STORAGE

If correctly used and maintained the device will have a minimum service life of 5 years. It is possible to store the device in suitable conditions for a maximum of 5 years. While in storage the battery needs to be recharged once per three months to avoid accidentally over-discharging the batteries. When using the device after storage, we recommend subjecting the device to a thorough maintenance check.

The device should be inspected by a qualified service engineer at least every 12 months for the following safety checks:

  • Any mechanical or functional damage on the device and accessories,

  • Performance of device in accordance to instructions for use,

  • The warning label’s legibility, and

  • The battery cycle count.