Instructions for Use

Technical specifications

Following are the technical information regarding the MESI SPO2 MODULE, its measurement and FingerClip Sensor (CS10299) provided within the package.

3.1 MESI PULSE OXIMETER UNIT

3.1.1 DIMENSIONS

Width: 40 mm (1.57 inches)

Depth: 48 mm (1.89 inches)

Height: 135 mm (5.31 inches)

Weight: 210 g

3.1.2 POWER & BATTERY

Power & Battery
Column AColumn B
Battery typeRechargeable Lithium-Polymer battery
Capacity1240 mAh
Examinations per battery charge> 8000
Continuous measurement56h
Charge time for depleted batteryapproximately 2 hours

3.1.3 CLASSIFICATION

Classification
Column AColumn B
Protection against electric shock Class II
Medical device classificationClass IIa
Applied partsType BF Applied part
Software classificationClass B
RF emissions (CIPSR 11)Group 1. Class B

3.1.4 OPERATING CONDITIONS

OPERATING CONDITIONS
Column AColumn B
Temperature, operating10° to 40°C
Relative humidity25 to 85% (no condensation)
Pressure during operation700 to 1060 hPa
Ingress protection ratingIP44

NOTE: The device is protected against the ingress of a solid object greater than 1 mm in size as well as against the harmful effects of spraying water from all sides.

⚠️If the device is used or stored outside the specified environmental parameters, the accuracy specified within the technical specifications of the device is not guaranteed. 

3.1.5 MEASUREMENT SPECIFICATIONS

Measurement specifications
Column AColumn B
SpO2 measurement range45 – 100 %
Pulse frequency range20 – 300 bpm
Plethysmogram0 – 28 LSB
Raw plethysmogram0 – 224 LSB
Signal quality0 – 100 %

NOTE: SPO2MD is calibrated to display functional oxygen saturation.

NOTE: For compliance with ISO 80601-2-61 the temperature has been measured on each SpO2 sensor with thermocouples (Type K ≤ 0,25 mm wire) procedure. During the measurement, a functional tester FLUKE ProSim 8 with artificial finger setting was used.

Data processing:
Column AColumn B
Data averagingnormal and 10 % VS
Update period5 samples per second

Alerts are implemented in MESI mTABLET UNIT and are being used for monitoring oxygen saturation and pulse rate value. It is not intended to be used for monitoring vital signs of the patient.

3.1.6 CONNECTIVITY

Data connectivity with MTABMD (Bluetooth 2.1 + EDR)

Connectivity
Receiving sectionColumn B
Frequency range2401.3 MHz – 2480.7 MHz
Bandwidth0.930 MHz
Connectivity
TransmitterColumn B
Output power0.5 – 4.5 dBm
Frequency range2401.3 MHz – 2480.7 MHz
ModulationGFSK

3.2 SpO₂ SENSORS

The SPO2MD is compatible with the following SpO₂ sensors:

  • FingerClip 1.2 m (CS10299)

  • FingerClip 2.5 m (CS10299-02)

  • EarClip sensor (CS10109)

  • SoftTip® Medium (CS10318)

  • SoftTip® Large (CS10319)

  • Wrap-Sensor (CS10329)

  • Y-Sensor (CS10532)

The SPO2MD package includes adult FingerClip Sensor (CS10299) for SpO₂ measurement.

Provided sensor:

SpO₂ FingerClip Sensor (CS10299)

Accuracy:

The accuracy for the approved sensor according to this standard is given as the root-mean-square difference (Arms) between the measured SpO₂ values and reference SaO₂ values. SpO₂ FingerClip (CS10299): 2,3 % (Arms)

Length of the sensor:

SpO₂ FingerClip Sensor (CS10299): 120 cm

Wavelength and Output Power:

RED: 660 nm at 3,5 – 4,5 mW

IR: 905 nm at 3,5 – 4,5 mW

This information is especially useful for clinicians performing photodynamic therapy.

NOTE: For more information regarding approved sensors and other accessories see instruction manual of the sensor available in the MESI package or contact your local distributor.