Instructions for Use

Maintenance

6.1 CHARGING THE BATTERY

The device can be only used on battery power therefore it has to be ensured that the battery is charged regularly. One battery charge is sufficient for approximately 8000 measurements or 56 hours of continuous use.

To charge the battery, place the MESI PULSE OXIMETER UNIT on the MESI LARGE CHARGING PLATE UNIT. During the charging process the light on the module will be flashing orange. Only when the device is completely charged it will start flashing green. For more information see chapter 5.5 MULTIFUNCTIONAL LED BUTTON.

If the capacity of the battery is significantly decreased after a certain period of intensive use, the battery is most likely spent, and it should be replaced. As the device does not contain any parts which can be replaced by a user, contact your local distributor or the manufacturer about replacing the battery.

6.2 CLEANING INSTRUCTIONS

⚠️Read and follow cleaning instructions carefully.

Reusable pulse oximetry sensors

The pulse oximeter sensor must be carefully cleaned and disinfected after every use and before use with a different patient. The sensor can be reused on the same patient, for the patient’s entire stay. If the sensor is covered with dust and dirt, it can affect the performance of the device. Observe the cleaning instructions in the Instructions for Use that came with the sensor.

⚠️Do not use aggressive cleaning agents, volatile liquids or excessive force when cleaning the device. Don’t dip the sensor in liquids or cleaning agents.

MESI PULSE OXIMETER UNIT

  • Before cleaning remove the MESI PULSE OXIMETER UNIT from the charging station and disconnect the sensor from the MESI PULSE OXIMETER UNIT.

  • Use a soft lint-free cloth dampened with a proper cleaning agent to clean and disinfect the exterior and interior of the unit/sensor.

  • With a dry cloth wipe off all residues of the cleaning solution.

  • Dry the equipment before next use.

Recommended cleaning agents:

  • Hydrogen peroxide (3%)

  • Ethanol (70%)

  • Isopropanol (70%)

  • Mild soap (diluted)

  • Ammonia (diluted)

  • Sodium hypochlorite bleach (diluted)

Not approved cleaning agents:

  • Organic solvents

  • Ammonia-based detergent

  • Abrasive cleaning agents

  • 100% alcohol, acetone, chlorine etc.

6.3 PRODUCT LIFE AND STORAGE

If correctly used and maintained the device will have a minimum service life of 5 years. It is possible to store the device in suitable conditions for a maximum of 5 years. While in storage the battery needs to be recharged once per month to avoid accidentally over-discharging the batteries. When using the device after storage, we recommend subjecting the device to a thorough maintenance check. The device should be inspected by a qualified service engineer at least every 12 months for the following safety checks:

  • any mechanical or functional damage on device and accessories,

  • performance of device in accordance to instructions for use,

  • legibility of labels,

  • battery cycle count.