Instructions for Use

Maintenance

7.1 Charging the battery

If you wish to use the device on battery power, you must ensure that the battery is charged regularly. One battery charge is sufficient for approximately 50 measurements.

If the capacity of the battery is significantly decreased after a certain period of intensive use, the battery is most likely spent and you should replace it. As the device does not contain any parts which can be replaced by a user, you should contact your dealer or the manufacturer about replacing the battery.

7.2 Cleaning instructions

It’s recommended to clean the device regularly with a soft, dry or damp cloth. Do not use aggresive cleaning agents, volatile liquids or excessive force when cleaning the device.

Do not wash the cuffs or immerse it in water. Also do not use petrol, thinners or similar solvents to clean the cuffs. To adequately clean the cuffs use a soft, moistened cloth and soap.

7.3 Recommendations for service and maintenance intervals

It is mandatory to calibrate the device once per year. Contact your dealer or the manufacturer about calibrating the device.

7.4. Product life and storage

If correctly used, maintained and regularly calibrated, the device will have a minimum service life of 5 years. You can store the device in suitable conditions for a maximum of 5 years.

When using the device after storage, we recommend you to subject the device to a thorough maintenance check and calibrate it.

7.5 Protecting the cuffs

In case of open wounds, protect the wounds with adequate impermeable dressing prior to applying the cuffs. Use of protective sleeves is recommended.

NOTE! Do not wash the cuffs in a washing machine or iron them.

NOTE! In the event of mechanical stress, the device must be calibrated!

NOTE! In the event of contact with infected body fluids, the device should be cleaned with non-aggressive cleaning agents and the cuffs should be immediately replaced with new. To avoid replacing the cuffs, use the cuff protection paper.

NOTE! It is mandatory to have the device calibrated once per year to ensure the correct functioning and accuracy. Contact your dealer or the manufacturer about calibrating the device.