Instructions for Use

Quick measuring guide

NOTE: Before using the device for the first time, read the instructions for use carefully and follow the recommendations and suggestions. This chapter only includes short instructions for the use of the MESI ABPI MD. For detailed descriptions of individual functions of the device, see chapter 5, Detailed instructions. Please keep for future reference.

NOTE: When using the device for taking the Ankle-Brachial Pressure Index measurement, the patient must lie down and remain still.

NOTE: When using the device for measuring the upper arm blood pressure, the patient must sit upright, keeping a straight back.

NOTE: The MESI ABPI MD is intended for use in medical institutions, where measurements must be carried out by adequately trained medical personnel. The MESI ABPI MD is not intended for home use.

NOTE: The MESI ABPI MD may be used on pregnant women.

NOTE: The MESI ABPI MD is not intended for use on newborns or children under the age of 10 years.

NOTE: In case of the presence of intravenous cannulas or arteriovenous (AV) fistulas, the cuffs and measurement can cause injury to the limb.

3 Quick guide

The MESI ABPI MD has two modes of operation:

The primary mode is for the measurement of the Ankle-Brachial Pressure Index, where you place the cuffs on the arm and the left and right ankles. The results of the measurement are the left Ankle-Brachial Pressure Index (LEFT ABI), the right Ankle-Brachial Pressure Index (RIGHT ABI) and the upper arm blood pressure (SYS, DIA).

The additional mode is for the independent measurement of the upper arm blood pressure. In this mode, you will only measure the upper arm blood pressure, so only place a cuff on the upper arm. The result of the measurement is the upper arm blood pressure (SYS, DIA).

Step 1

Connect the MESI ABPI MD to the mains electricity using the AC/DC power supply. The power socket is located at the back of the device and is labelled “5V DC”. Connect the cuffs to the device by inserting the tube of each cuff into the connector of the same colour on the device.

ABPI MD quick guide - step 1

Step 2

Turn the device on using the ON/OFF button . The display will show the home screen. It is not necessary to change any settings for the basic functioning of the device. If you are using the MESI ABPI MD device for the first time, the language, time and date menu will appear. To correctly set the menu, see section 5.1.

ABPI MD quick guide - step 2

3. 1 Ankle-Brachial Pressure Index measurement

Step 3

The patient must lie down and remain still. Place the cuffs on the upper arm (cuff labeled “ARM”), the lower left leg (cuff labeled “LEFT ANKLE”) and the lower right leg (cuff labeled “RIGHT ANKLE”). Observe the following colour markings:

🟥 RED > left or right upper arm

🟨 YELLOW > lower left leg

🟩 GREEN > lower right leg

ABPI MD quick guide step 3

Place the cuffs so that there is a finger’s width of room between the limb and the cuff.

ABPI MD cuff placement

The cuff on the upper arm must be placed so that the artery label (ARTERY) points towards the elbow. When placing the cuff, observe the image below. The cuff tube should point towards the fingers.

ABPI MD cuff placement

The ankle cuffs must be placed so that the ankle label (MEDIAL ANKLE) points towards the inner side of the ankle. When placing the cuff, observe the image below.

ABPI MD ankle cuff placement

Step 4

The patient must lie still in the supine position. When you are ready, push the START button. The measurement process will begin. The patient must not move during the measurement process. The entire process takes approximately 1 minute.

During the measurement process, the screen will display the current arterial pressures and the arterial pressure waveforms.

ABPI MD quick guide
ABPI MD quick guide
ABPI MD quick guide
ABPI MD quick guide

Step 5

When the measurement process is completed, the screen will display the results. The Ankle-Brachial Pressure Index is displayed in green, yellow or red, depending on the value. The colour scale is taken from the “Guidelines for the Management of Patients With Peripheral Artery Disease” published by the Journal of the American College of Cardiology.

ABPI MD results

Step 6

Return to the home screen using the HOME button. From here, you can view the measurement history and change the settings. For an additional description of the specifications of the MESI ABPI MD and detailed instructions for its use, see the remaining chapters of the instructions for use.

ABPI MD home

3.2 Upper arm blood pressure measurement

Step 3

The patient must sit comfortably on a chair. They must sit upright, keeping a straight back. Only place the red upper arm cuff (cuff labeled “ARM”) on the patient). In this mode of operation, the green and yellow cuffs are not placed on the patient’s limbs.

ABPI MD blood pressure

Place the cuff so that there is a finger’s width of room between the limb and the cuff.

The cuff on the upper arm must be placed so that the artery label (ARTERY) points towards the elbow. When placing the cuff, observe the image below. The cuff tube should point towards the fingers.

The lower arm should lie freely on a flat surface. The cuff should be level with the patient’s heart.

ABPI MD blood pressure measurement

Step 4

Push the BLOOD PRESSURE MEASUREMENT button to change the mode to independent blood pressure measurement.

When you are ready, push the START button. The measurement process will begin. The patient must not move during the measurement process. The entire process takes approximately 1 minute. During the measurement process, the screen will display the current arterial pressure and the arterial pressure waveform.

ABPI MD blood pressure

Step 5

When the measurement process is completed, the screen will display the results of the upper arm blood pressure measurement.

ABPI MD BP results