Instructions for Use

Errors

Errors
ErrorDescriptionSolution
E3: Inflation insufficiency. Check the placement of the cuffs and repeat the measurement.The cuff was insufficiently inflated.Check the placement of the cuffs and repeat the measurement.
E7: Deflation insufficiency. Check the placement of the cuffs and repeat the measurement.The cuff was insufficiently deflated.The cuff is not attached or is not attached correctly. Attach the cuff correctly and repeat the measurement.
E8: An error occurred during systolic pressure calculation.An error occurred during systolic pressure calculation.Repeat the measurement. if the error is repeated, the measured value is outside the measurement range of the device.
E9: An error occurred during diastolic pressure calculation.An error occurred during diastolic pressure calculation.Repeat the measurement. if the error is repeated, the measured value is outside the measurement range of the device.
E10: An error occurred during mean pressure calculation.An error occurred during mean pressure calculation.Repeat the measurement. if the error is repeated, the measured value is outside the measurement range of the device.
E11: A large pressure fluctuation was detected. The patient may have moved.A large pressure fluctuation was detected. The patient may have moved.The results may be incorrect. Repeat the measurement.
N21: MESI TUBLESS CUFF Module not in range.MESI TUBELESS CUFF UNIT is out of range (timeout).Bring the MESI TUBELESS CUFF UNIT closer to the tablet and restart the ABI application.
N21: MESI TUBLESS CUFF Module not in range.Connection between MESI TUBELESS CUFF UNIT and MESI mTABLET UNIT is not optimal.Remove any mobile devices in proximity of MESI TUBELESS CUFF UNIT or bring the MESI mTABLET UNIT closer to the MESI TUBELESS CUFF UNIT.
N21: MESI TUBLESS CUFF Module not in range.Depleted battery.Place the MESI TUBELESS CUFF UNIT on the charging station.
N22: Module not paired with MESI mTABLET UNIT.MESI TUBELESS CUFF UNIT is not paired to the MESI mTABLET UNIT.Pair the MESI TUBELESS CUFF UNIT to the MESI mTABLET UNIT. For pairing see Chapter 5.1.5 Pairing.
N41: No cuff connected to MESI TUBLESS CUFF UNIT.The cuff is disconnected from the MESI TUBELESS CUFF UNIT.Disconnect the cuff from the MESI TUBELESS CUFF UNIT, turn it upside-down and reconnect it to the device.
E42: Please check the cuff position and repeat the measurement. If error is repeated, the measured value is outside the measurement range of the device.The cuff on the patient limb is not correctly positioned.Follow chapter 5.3.1 Cuff placement and place the cuffs correctly.

NOTE: For “Abnormally weak pulse detected” message, please see chapter 8 ERRORS.